• fukuda denshi ds 7100 service manual

    fukuda denshi ds 7100 service manual

    Download Link: ➡ fukuda denshi ds 7100 service manual



    File Name: fukuda denshi ds 7100 service manual.pdf
    Size: 2733 KB
    Type: PDF, ePub, eBook
    Category: Book
    Uploaded: 15 May 2019, 18:36 PM
    Rating: 4.6/5 from 684 votes.


    Last checked: 4 Minutes ago!









    ⬆ LINK ⬆




















    In order to read or download fukuda denshi ds 7100 service manual ebook, you need to create a FREE account.

    eBook includes PDF, ePub and Kindle version



    ✔ Register a free 1 month Trial Account.
    ✔ Download as many books as you like (Personal use)
    ✔ Cancel the membership at any time if not satisfied.
    ✔ Join Over 80000 Happy Readers


    Book Descriptions:

    fukuda denshi ds 7100 service manual

    System Configuration Ward Setup ” for user key setup. 2. Clean the touch panel. If “ LEAD OFF ” or other message is displayed, the key lock message will not be displayed. 3. Wipe the touch panel using c leaning cloth. 4. Press again the Key Loc k key for more than 2 seconds. The message will disappear and the keys will be activate again. CAUTION ? Do not clean the touch panel using strong acid. ? A special coating is applied to the surface of the touch panel. Wipe the surface with the soft cleaning cloth provided as optional accessory or with commercially available eyeglass cleaning cloth. CAUTION ? Clean the equipment frequently so stains can be removed easily. ? To prevent injury, it is recommended to wear gloves when cleaning the equipment. ? Do not allow liquids or cleaning solution to enter the monitor or connectors. ? Do not use organic solvents, thinner, toluene and benzene to avoid damaging the resin case. ? Do not polish the housing with abrasive or chemical cleaner. ? When sterilizing the entire room using a spray solution, pay close attention not to have liquids get into the monitor or connectors. ? Use only neutral detergent to clean the housing. Do not use chemical cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent and chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent for house and furniture), or sharp-edged tools. The surface resin coating may be damaged, resulting in discoloration, scratches, and other problems. Disinfecting the Blood Pressure Transducers Disinfect the blood pressure transducers according to the manufacturer’s guidelines. Disinfecting the Temperature Probe s Disinfect the Temperature Probe s according to the manufacturer’s guidelines. Cleaning and D isinfecting the SpO 2 Transducer. Do not soak the sensor in water or antiseptic solution. ? Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not disinfect by applying radioactive rays, steam, or ethylene oxide. ?

    • fukuda denshi ds 7100 service manual, fukuda denshi ds 7100 service manual pdf, fukuda denshi ds 7100 service manual.

    Handling After Use. Do not apply excessive force when disconnecting the cables. Always pull on the connector housing and not on the c able. ? Clean the unit, accessories, and cables, and keep them together in one place for next use. ? Always check for adequate supply of disposable accessories such as ECG electrodes. If any shortage, contact our service representative and supply as necessary. Handling the Display Panel. The display panel utilizes exclusive fluorescent light for the backlight. As this fluorescent light tube has product life cycle, it needs to be replaced periodically. If the display becomes dark, scintillates, or does not light, contact your nearest service representative. ? The LCD used for the display panel utilizes highly accurate picture element s of pixel s over 99.99%, but there may be an absence ( less than 0.01% ) or constant lighting of pixel s. Storing the Device. Store in a place where the device will not be exposed to splashing water. ? Store in a place where the device will not be adversely affected by atmospheric pressure, temperature, humidity, ventilation, sunlight, dust or atmosphere containing salt or sulfur. ? Store in a level area where the device is not exposed to vibration and shock (inclu d ing during transportation). ? The following environmental conditions should be observed when storing the device. Storage Temperature: ? 10 ? 60 ? C Storage Humidity: 10 ? 95% (at 60 ? C) Storing the Recording Paper The DS-7100 system utilizes heat sensitive recording paper. If placed in a high temperature for long period of time, the print may become indistinct, and unable to read.Cleaning the Displa y Panel Since this device incorporates a touch screen, finger prints and other stains are likely to appear on the display panel. Follow the procedure below to clean the display panel. 1. Press the Key Loc k key for more than 2 seconds. Reference The Key Lock key needs to be preprogrammed as user key. Refer to ” 8.

    When reusing the device which was left unused for a while, always check that the device operates properly and safely before use. To ensure safety, reliability, and high performance, a “ Daily Check ” and “ Periodic Inspection ” must be performed. We are not liable for any accident arising from lack of maint e nance. CAUTION ? Do not open the housing of this device. ? Avoid alcohol or other liquids from getting into the equipment. ? Daily Check Perform daily inspection using the “ Daily Check List ” on the next page. ? Periodic Check Periodic inspection of medical electronic equipment is mandatory to prevent failures and accidents and to ensure safety and reliability. Periodic maintenance may be performed by each medical institution or by a third party by concluding a “ Maintenance Contract ”. For more details, contact your nearest service representative. Time-Change Components To ensure reliability of safety, function, and performance of this device, the time-change components must be replaced periodically. When replacing, contact our service representative. CAUTION The time-change components must be replaced at specified period. 10 Maintenance Check Inspected Date Inspected by Location Device Type Serial No. Date of Purchase Item Details Criteria Judgement Appearance Visually check the exterior for scratches, cracks, deformation, and rust. The environmental condition (ex.ECG The “ LEAD OFF ” message is displayed. Cause: The electrode is detached, or is not making good electrical contact with the skin. Solution: ? Check if the electrodes are properly attached. ? Replace the electrode, or check the lead cable. The “ ECG failed ” message is displayed. Note: Using 4-electrode or 5-electrode instead of 3-electrode allows more accurate QRS detection. Cause 2: The electrode contact is poor. Electrical blanket or other noise source is near the patient. Solution: Attach the electrode firmly. ? Replace the lead cable if defective. ?

    OXISENSOR is a disposable sensor. Do not reuse or attempt re sterilization. Cleaning and D isinfecting the NIBP C uff. Do not soak the sensor in water or antiseptic solution. ? Wipe the NIBP cuff with disinfectant such as 70% alcohol. Do not disinfect by applying radioactive rays, steam, or ethylene oxide. Cleaning and D isinfecting the ECG lead Cable. Do not soak the sensor in water or antiseptic solution. ? Wipe the ECG lead cable with disinfectant such as 70% alcohol. Do not disinfect by applying radioactive rays, steam, or ethylene oxide. 10 Cleaning 10 Cleaning 10 Cleaning Handling the Battery. The battery pack can be continually used for more than 300 times (or about 1 year) under normal temperature, but the continuous use will degrade the battery and shortens the usable time. ? When the battery operation time becomes short even after it is fully charged, the battery pack needs to be replaced. ? When the charge time of the battery pack becomes short, the battery pack needs to be replaced. ? When the battery pack level becomes low, charge the battery well in advance for the next use. Storing the Battery To take advantage of the characteristic of battery pack, pay attention to the following when storing. Storage Temperature and Humidity. Store in an environment specified below without corrosive gas. Storage Period Storage Temperature Storage Humidity Within 30days ? 20 ? 60 ? C 30 days ? 90 days ? 20 ? 45 ? C 90 days ? 1year ? 20 ? 35 ? C 65 ? 20% ? Do not store in an environment outside the specified temperature range or excessive high humidity. Long-term Storage ? If left installed in the monitor for long period of time, the electrolyte may leak, or inactivate the battery which degrades the capacity recovery after storage. Therefore, always remove the battery from the monitor when storing for long period of time. Contact our service representative when removing the battery. About the Maintenance Check Periodic inspection must be performed.

    The “ Pacemaker error ” message is displayed. Cause: The pacemaker pulse is detected 16 pulses or more per second. Solution 1: Attach the electrodes firmly. ? Replace the lead cable if defective. ? If any noise source is near the patient, locate it away from the patient as much as possible. The “ ECG not connected ” message is displayed. Cause: When the ECG relay cable is disconnected during ECG monitoring, this message will be displayed. Solution 1: To cease monitoring, press the Alarm Silenc e key to clear the message and silence the alarm. Solution 2: To conti nue monitoring, plug in the ECG relay cable. This will clear the message and silence the alarm. Cause: Heartbeat is interfering and superimposed on the respiration waveform. Solution: Place the ele ctrode as shown below where the heartbeat will be less likely to interfere. “ 0 ” is displayed for respiration rate, or apnea alarm is generated. Cause: The respiration waveform amplitude is below the detection level ( 0.2O ). Solution: Change the electrode site. The respiration waveform and respiration rate is not displayed. Cause 1: The ECG relay cable designed for electrosurgical knife is used. Solution: The impedance respiration can not be measured if the cable design ed for electrosurgical knife is used. Use the standard ECG relay cable if not using the electrosurgical knife. Cause 2: The impedance respiration measurement is ceased. Note: If the pacemaker with the minute ventilation measuring function is used, turn OFF the impedance respiration measurement. Otherwise, both the pacemaker and the monitor will not be able to perform accurate measurement. The measured data is displayed as “ ??? ”. Cause: The respiration rate is outside the measurement range. Invasive Blood Pressure The “ BP1 Transducer OFF ”, “ BP2 Transducer OFF ” message is displayed. Cause: The transducer for BP1 or BP2 is not connected. Solution: Connect the transducer.

    If any noise source is near the patient, locate it away from the patient as much as possible. ECG waveform contains noise. The “ Artifact ” message is displayed. Cause 1: The electrode contact is poor. Solution: Attach the electrodes firmly. ? Replace the lead cable if defective. ? If any noise source is near the patient, locate it away from the patient as much as possible. Cause 2: EMG is interfering. Solution: ? Change the electrode site to a location where EMG will less likely to interfere. ? Select ESIS mode for the filter mode. Note: Selecting a ESIS mode for the filter mode will decrease the QRS amplitude and may result in not counting the heart rate. The “ Check electrode ” message is displayed. Cause: The electrode contact with the skin is poor. There is substantial contact resistance between the electrodes. Solution: Replace all the electrodes. Use the electrodes of the same type. The “ ECG unit error ” message is displayed. Cause: A communication error with the ECG measuring unit exists. Solution: The breakage of wire or failure of the ECG unit can be considered. Contact our service representative. 10 Troubleshooting Solution: ? Check the electrode application. ? Replace the electrode, or check the lead cable. Heart rate is not counted. Heart rate is low. Cause: The ECG waveform amplitude is below the QRS d etection level (0.3mV). Solution: Change the electrode site, or select a lead with higher QRS amplitude. Also, if large amount of noise is interfering, the noise may be erroneously detected as QRS. It is recommended to change the electrode site and increase the ECG amplitude. Heart rate is not counted, and “LEAD OFF” message is displayed. Cause: The electrode of the displayed lead type is detached, or is not making good electrical contact with the skin. Solution: ? Check the electrode application. ? Replace the electrode, or check the lead cable. Artificial pacemaker is not displayed. Solution: Select Used for the pacemaker use.

    The monitor repeats the measurement, or “ ? ? ? ” is displayed for the numeric data. Cause 1: The measurement accuracy is not reliable due to body motion artifact. Solution: Have the patient stay still as much as possible during the measurement. Cause 2: The pulse is too small to acquire reliab le measurement accuracy. Solution: Check if the cuff application is proper, and if the cuff size is corresponded to the selected patient type. The “ Check NIBP hose ” message is displayed. Cause: The applied pressure to the cuff has exceeded the maximum limit. The measurement time has exceeded the maximum limit. Solution: Check if the cuff application is proper, if the cuff size is corresponded to the selected patient type, or if the air hose is not bent. After checking the above, perform the measurement again. If the same message is displayed again, a failure of the equipment can be considered. Cease the measurement, and contact our service representative. 10 Troubleshooting Cause: The zero balancing before the measuremen t has failed, and measurement could not be started. Solution: The body movement or other artifact may cause zero balance failure. During the measurement, have the patient stay still as much as possible. If the same message is displayed again, the failure of the equipment can be considered. Cease the measurement, and contact our service representative. The time of measurement disappears and the numeric data is displayed as “ ? ? ? ”. Cause: The NIBP data will be erased when the preprogrammed NIBP erase ti me has elapsed. Solution: Select the appropriate time for NIBP data erase time from 10min, 30min, 60min, 24hrs which best fits the monitoring purpose. Temperature The “ Wrong Temp Probe ” message is displayed. Cause 1: The YSI-700 is used. Solution: Use the YSI-400 temperature probe for measurement. The YSI-700 can not be used with the DS-7100 series. Cause 2: There is a contact failure of the temperature probe.

    Solution: Check if the temperature probe is properly inserted. The numeric data is displayed as “ ??? ”. Cause: The temperature measurement is outside the measurement range. The “ TEMP not connected ” message is displayed. Cause: When the temperature sensor is disconnected during temperature monitoring, this message will be displayed. Solution 2: To continue monitoring, plug in the temperature sensor. This will clear the m essage and silence the alarm. The “ TEMP auto check ” message is displayed. The numeric data is displayed as “ ? ? ? ”. Cause: The temperature is calibrated once every hour on this monitor. During calibration, the numeric data will be displayed as “ ? ? ? ”. Solution: The calibration will complete in 10 seconds. If the calibration does not complete within 10 seconds, cease the measurement and contact our service representative. The “ TEMP unit check ” message is displayed. Cause: Error is detected during te mperature calibration. Solution: A unit failure can be considered. Cease the measurement and contact our service representative. Cause: The sampling tube is clogged. Solution: Replace t he sampling tube. The “ Self-diag CO 2 ” message remains displayed. Cause: An error has occurred to the self-check procedure at power ON. Solution: The CO 2 unit failure can be considered. The “ Initializing CO 2 ” message does not disappear. Cause: An err or has occurred during the initialization at power ON. The “ Check CO 2 unit ” message is displayed. Cause 1: The exhaust connector is clogged. Solution: After checking the exhaust system and removing the clog, press the “ Restart CO 2 ” key on the CO 2 configuration menu. Cause 2: The sampling tube or nasal prong is clogged. Solution: After checking the inhalation system and removing the clog, press the “ Restart CO 2 ” key on the CO 2 configuration menu. Cause 3: The CO 2 unit needs to be replaced. Solution: Contact our service representative. The “ CO 2 unit error ” message is displayed.

    The “ BP1 not zero balanced ”, “ BP2 not zero balanced ” message is displayed. Cause: The BP zero balance has not been performed since the power is turned ON. Solution: Open the three-way cock of the transducer to air and perform zero balan ce. The measured data is displayed as “ ? ? ? ”. Cause: The BP zero balance has not been performed since the power is turned ON. Solution: Open the three-way cock of the transducer to air and perform zero balance. 10 Troubleshooting Red White Black Cause: Blood pressure line has not been zero balanced. Solution: Open the three-way cock of the transducer to air and perform zero balance. The measured data is displayed as “ ??? ”. Cause: The BP value is outside the measurement range. Solution: Perform zero balance again. The “ BP not connected ” message is displayed. Cause: When the BP interface cable or 2ch BP conversion cable is disconnected during BP monitoring, this message will be displayed. Solution 2: To continue monitoring, plug in the BP interface cable or 2ch BP conversion cable. This will clear the message and silence the alarm. The “ Incorr. BP cable ” message is displayed. Cause: The cable other than 2ch BP conversion cable is plugged in to the BP connector. Solution: Use the 2ch BP conversion cable. SpO 2 The “ Check SpO 2 sensor ” message is displayed. Cause: Sensor is detached from the patient. Solution 1: Check if t he sensor part is properly attached to the patient. Solution 2: Check if the light emitting part and light receiving part of the sensor LED is aligned. The “ Pulse search ” message is displayed. Cause: The amplitude of the pulse waveform is low, or the s ensor is not positioned correctly. Solution: Check if the light emitting part and light receiving part of the sensor LED is aligned. The “ No pulse detect ” message is displayed. Cause: The amplitude of the pulse waveform is low, or the sensor is not pos itioned correctly. The “ Motion Artifact ” message is displayed.

    Cause: There is excessive body motion of the patient. Solution: Change the sensor position where the body motion will have less effect. The pulse waveform is not displayed, or interrupted Situation: “ Check SpO 2 sensor ” is displayed. Cause 1: The amplitude of the pulse waveform is low, or the sensor is not positioned correctly. Solution: Ch eck if the light emitting part and light receiving part of the sensor LED is aligned. Cause 2: Sensor is defective. Solution: Replace the sensor. Cause 3: SpO 2 sensor is not firmly connected to the SpO 2 input connector. Solution: Make sure the SpO 2 sensor is securely connected. Cause 4: Sensor is exposed to light. Solution: Place a black or dark cloth over the sensor to avoid direct sunlight. Also when not used, avoid placing the sensor in light or unplug the sensor from the connector. Cause: There is excessive body motion of the patient which disables correct measurement. Solution: 1. Have the patient lie still as much as possible. 2. Relocate the sensor, or change the sensor to which the body motion will ha ve less influence. The “ SpO 2 unit error ” message is displayed. Cause 1: There is a failure of communication with the SpO 2 measurement unit. Solution: Breaking of wire or SpO 2 unit failure can be considered. Contact our service representative. The “ SpO 2 sensor fault ” message is displayed. Cause: Sensor is defective. The “ SpO 2 not connected ” message is displayed. Cause: When the SpO 2 relay cable is disconnected during SpO 2 monitoring, this message will be displayed. Solution 2: To continue monitoring, plug in the SpO 2 relay cable. Non-Invasive Blood Pressure The cuff is not inflated although the pump is operating. Cause1: The air hose is not firmly connected, and the air is leaking. Solution: Check if the air hose is properly connected. Cause 2: The cuff size is not corresponded to the selected patient type. Solution: Check if the cuff size is corresponded to the selected patient type.

    Cause: There is a communication error with the CO 2 unit. Solution: The break of wire or CO 2 unit failure can be considered. There is substantial measurement error. Cause 1: 20 minutes have not yet elapsed since the power is turned ON. Solution: For 20 minutes from turning ON the power, there will be a substantial measurement error. Cause 2: The calibration is not properly performed. Solution: Perform CO 2 calibration again. The “ CO 2 not connected ” message is displayed. Cause: When the filter line is disconnected during CO 2 monitoring, this message will be displayed. Solution 2: To continue monitoring, plug in the filter line. This will clear the message and silence the alarm. 10 Troubleshooting Situation: The “ Paper Out ” message is displayed on the upper left of the screen. The “ Paper Out ” message is displayed on the Rec. Solution: Install a new pad of paper into the pap er magazine. Situation: The “ Magazine Open ” message is displayed. Cause: The paper magazine is open. Solution: Close the magazine. Situation: The “ Paper jammed ” message is displayed. Cause: The paper is jammed. Solution: Open the magazine and ins tall the paper correctly. Situation: No message is displayed, but recording can not be performed. Cause: The recording paper is not correctly installed. The front and backside of the paper is set oppositely. Solution: The “ END ” printed side of the paper should be facing down in the magazine. The second waveform and third waveform are not recorded. Situation: The second waveform and third waveform are not recorded for manual recording or alarm recording. Cause: The second waveform and third waveform are not set on the recording setup menu. Solution: Set the second waveform and the third waveform on each recording setup menu. The “ Recorder error ” message is displayed. Cause: The thermal head temperature has increased. Solution: A damage to the the rmal head can be considered. Telemetry The “ Telemetry unit error ” message is displayed.

    Cause: There is a communication error with the telemetry transmission unit. Solution: The breaking of wire or telemetry transmiss ion unit failure can be considered. There is no reception at the telemetry center. Cause: The channel ID or group ID is not corresponded with the telemetry receiver. Solution: Set the correct channel ID and group ID. The BP waveform of 100mmHg or above can not be properly received. Cause: The BP waveform and scale is not corresponded. Solution: When BP waveform is above 100mmHg, set the BP scale above 100mmHg. Cause: A system error has occurred. Solution: Turn off the power, unplug the power cable, and contact our service representative. The “ Adjusting ” message is displayed. Numbers are displayed large on the display. Cause: This is the test mode. Stop using the device immediately. Solution: Restart the system. The test mode will be cancelled. If the same situation is observed again, contact our service representative. Turn off the DIP switch No.1. The data is initialized each time the power is turned ON. Cause 1: The internal switch is set to initialize. Solution: The internal switch setting needs to be changed. Set the rotary switch to 0. Cause 2: The battery for backup memory is depleted. Solution: The battery needs to be replaced. The display is not clear. Cause 1: The display brightness is not adjusted. Solution: Due to the LCD display characteristic, the visible range is limited. Adjust to the appropriate brightness. Cause 2: The monitor is set to the night mode. Solution: Cancel the night mode. The system does not start although the power switch is turned ON. Cause 1: The power cable is not connected. The battery is not charged. Solution: Tu rn off the power and connect the power cable. If the battery is not charged, use the power cable until the battery charging is complete. Cause 2: Incorrect IC card is inserted. Solution: Turn off the DIP switch No.8. The clock is often delayed.

    Cause: The battery for the backup memory is depleted. Check if the time is delayed when the power is turned off. Contact our service representative. 10 Troubleshooting Cause 1: The battery life has expired. Solution: The battery pack is a consumable product. Replace it once a year. Cause 2: The ambient temperature is too high or too low. Solution: For safety, the charging operation will be in a standby mode w hen the battery pack temperature becomes excessively high or low. The charging will automatically resume when appropriate temperature is reached. Charge the battery in an ambient temperature of 10 ? 30 ? C. The charge lamp on the patient monitor does not light. Cause 1: The AC power cable is disconnected. Solution: Plug in the AC power cable. The battery pack can be charged only during the AC operation. Cause 2: The battery pack is not installed. Solution: The battery pack is optional. If a battery pack is required, contact our service representative and install the battery pack. Cause 3: The battery life has expired. During the charging procedure, the charge lamp (orange) does not switch to charge complete status (green) and extinguishes. Cause 1: The battery pack temperature is too high or too low. Solution: For safety, the charging operation will be in a standby mode when the battery pack temperature becomes excessively high or low. Cause 2: The breakdown of battery pack can be considered. Solution: If the charging operation does no t complete within the specified charging time, the charging operation will cease for safety purpose. Contact our service representative and replace the battery pack. Solution: The battery pack is a consumable produ ct. The “ Charge battery ” message is displayed. Cause: The AC power cable is disconnected. The battery pack can be charged only during the AC operation. This section states the specification and performance of this equipment. Part 1: General requir e ments for safety ? 2.

    Collateral standard: Electromagnetic compatibility.Power Consumption 82 VA 60W RF O utput P ower: -15 dBm Standard, 0dBm MAX Channel Spacing: 12.5 kHz Modulation Mode: Digital, F requency S hift K eying (FSK) Backup Item “ 0 ”: Setup item will be retained even when the power is turned OFF. “ ? ”: Setup item will be retained even when the power is turned OFF. When discharging procedure is performed, the value will be reset to initial setting.


  • Commentaires

    Aucun commentaire pour le moment

    Suivre le flux RSS des commentaires


    Ajouter un commentaire

    Nom / Pseudo :

    E-mail (facultatif) :

    Site Web (facultatif) :

    Commentaire :